User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 83: (1 year) CAL VER DATE: 26-Aug-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 61 NUMBER OF LINES: 217 CONFIGURATION: Fluke 5700A CONFIGURATION: Fluke 5725A ============================================================================= # # Source: # Fluke 80 Series Service Manual. # (PN 834168 April 1989, Rev. 3, 7/93, Change/Errata 8/94). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # Sub Fluke 83,85,87 Capacitance (1 year) CAL VER # # Required Files: # 83_85dsp.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5700A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5700A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 RSLT =DISPLAY TEST 1.006 PICE 83_85dsp 2.001 DISP No warm-up time requirement. 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 90%. 2.002 ASK- U 2.003 HEAD {* ROTARY SWITCH TESTS *} 2.004 JMP 3.001 2.005 EVAL 3.001 DISP Hold down the RANGE button while turning the UUT rotary 3.001 DISP function switch from OFF to VDC. Release the RANGE 3.001 DISP button and turn the switch to {* VAC *}. 3.002 MATH MEM1 = 0 3.003 MEMI Enter UUT reading: 3.004 MEME 3.005 MEMC display 12U 4.001 DISP Rotate the UUT function switch to {* VDC *}. 4.002 MATH MEM1 = -32 4.003 MEMI Enter UUT reading: 4.004 MEME 4.005 MEMC display 12U 5.001 DISP Rotate the UUT function switch to {* mVDC *}. 5.002 MATH MEM1 = -64 5.003 MEMI Enter UUT reading: 5.004 MEME 5.005 MEMC display 12U 6.001 DISP Rotate the UUT function switch to {* Ohms *}. 6.002 MATH MEM1 = -96 6.003 MEMI Enter UUT reading: 6.004 MEME 6.005 MEMC display 12U 7.001 DISP Rotate the UUT function switch to {* Diode *}. 7.002 MATH MEM1 = -128 7.003 MEMI Enter UUT reading: 7.004 MEME 7.005 MEMC display 12U 8.001 DISP Rotate the UUT function switch to {* mA/A *}. 8.002 MATH MEM1 = -160 8.003 MEMI Enter UUT reading: 8.004 MEME 8.005 MEMC display 12U 9.001 DISP Rotate the UUT function switch to {* uA *}. 9.002 MATH MEM1 = -192 9.003 MEMI Enter UUT reading: 9.004 MEME 9.005 MEMC display 12U 10.001 ASK+ U 10.002 HEAD {* AC VOLTAGE TESTS *} 10.003 JMP 11.001 10.004 EVAL 11.001 DISP Rotate the UUT function switch to OFF then VAC. 11.002 DISP Connect the 5700A and the UUT as follows. 11.002 DISP [32] 5700A OUTPUT HI to UUT V Ohm Diode 11.002 DISP [32] 5700A OUTPUT LO to UUT COM 11.002 DISP [32] 5700A SENSE HI to UUT V Ohm Diode 11.002 DISP [32] 5700A SENSE LO to UUT COM 11.003 HEAD AC VOLTAGE TESTS: { 400mV Range } 11.004 5700 400 350.0mV 3.9U 60H 2W 12.001 5700 400 350.0mV 5.7U 1kH 2W 13.001 5700 400 350.0mV 7.4U 5kH 2W 14.001 HEAD AC VOLTAGE TESTS: { 4V Range } 14.002 5700 4 3.500V 0.038U 60H 2W 15.001 5700 4 3.500V 0.056U 1kH 2W 16.001 5700 4 3.500V 0.073U 5kH 2W 17.001 HEAD AC VOLTAGE TESTS: { 40V Range } 17.002 5700 40 35.00V 0.38U 60H 2W 18.001 5700 40 35.00V 0.56U 1kH 2W 19.001 5700 40 35.00V 0.73U 5kH 2W 20.001 HEAD AC VOLTAGE TESTS: { 400V Range } 20.002 5700 400 350.0V 3.8U 60H B1 2W 21.001 5700 400 350.0V 5.6U 1kH B1 2W 22.001 5700 400 350.0V 7.3U 5kH B1 2W 23.001 HEAD AC VOLTAGE TESTS: { 1000V Range } 23.002 5700 1000 900V 12U 60H B1 2W 24.001 5700 1000 900V 26U 1kH B1 2W 25.001 5700 1000 900V 26U 5kH B1 2W 26.001 DISP Disconnect the UUT from the 5700A. 26.002 CALL Sub Fluke 83,85,87 Freq & Duty Cycle (1 year) CAL VER 26.003 HEAD {* DC VOLTAGE TESTS *} 26.004 JMP 27.001 26.005 EVAL 27.001 DISP Press the Hz button twice to select VDC. 27.001 DISP 27.001 DISP Press the RANGE for 2s to select autorange. 27.002 HEAD DC VOLTAGE TESTS: { 4V Range } 27.003 5700 4 3.500V 0.012U 2W 28.001 HEAD DC VOLTAGE TESTS: { 40V Range } 28.002 5700 40 35.00V 0.12U 2W 29.001 5700 40 -35.00V 0.12U 2W 30.001 HEAD DC VOLTAGE TESTS: { 400V Range } 30.002 5700 400 350.0V 1.2U 2W 31.001 HEAD DC VOLTAGE TESTS: { 1000V Range } 31.002 5700 1000 1000V 4U 2W 32.001 HEAD {* DC MILLIVOLT TEST *} 32.002 DISP Rotate the UUT function switch to mVDC. 32.003 HEAD DC MILLIVOLT TEST: { 400mV Range } 32.004 5700 400 350.0mV 1.2U 2W 33.001 HEAD {* RESISTANCE TESTS *} 33.002 DISP Rotate the UUT function switch to Ohms. 33.003 HEAD RESISTANCE TESTS: { 400 Ohm Range } 33.004 5700 400 190.0Z 0.9U CW 34.001 HEAD RESISTANCE TESTS: { 40 kOhm Range } 34.002 5700 40 19.00kZ 0.09U CW 35.001 HEAD RESISTANCE TESTS: { 4 MOhm Range } 35.002 5700 4 1.900MZ 0.009U 2W 36.001 HEAD RESISTANCE TESTS: { 40 MOhm Range } 36.002 5700 40 19.00MZ 0.22U 2W 37.001 HEAD {* CONDUCTANCE TEST *} 37.002 DISP Press the Range button twice for 40nS range. 37.003 HEAD CONDUCTANCE TEST: { 40nS Range } 37.004 5700 40 10.00nY 0.20U 2W 38.001 HEAD {* DIODE TEST *} 38.002 DISP Rotate the UUT function switch to Diode. 38.003 5700 3.000V 0.061U 2W 39.001 CALL Sub Fluke 83,85,87 Capacitance (1 year) CAL VER 39.002 HEAD {* DC MILLIAMP TESTS *} 39.003 JMP 40.001 39.004 EVAL 40.001 DISP Rotate the UUT function switch to mA/A. 40.002 DISP Connect the 5700A and the UUT as follows. 40.002 DISP [32] 5700A OUTPUT HI to UUT mA/uA 40.002 DISP [32] 5700A OUTPUT LO to UUT COM 40.003 HEAD DC MILLIAMP TESTS: { 40mA Range } 40.004 5700 40 35.00mA 0.16U 2W 41.001 HEAD DC MILLIAMP TESTS: { 400mA Range } 41.002 5700 400 350.0mA 1.6U 2W 42.001 HEAD {* AC MILLIAMP TESTS *} 42.002 DISP Press the BLUE button to select AC. 42.003 HEAD AC MILLIAMP TESTS: { 40mA Range } 42.004 5700 40 35.00mA 0.44U 60H 2W 43.001 5700 40 35.00mA 0.44U 1kH 2W 44.001 HEAD AC MILLIAMP TESTS: { 400mA Range } 44.002 5700 400 350.0mA 4.4U 60H 2W 45.001 5700 400 350.0mA 4.4U 1kH 2W 46.001 HEAD {* DC MICROAMP TESTS *} 46.002 DISP Rotate the UUT function switch to uA. 46.003 HEAD DC MICROAMP TESTS: { 400uA Range } 46.004 5700 400 350.0uA 1.7U 2W 47.001 HEAD DC MICROAMP TESTS: { 4000uA Range } 47.002 5700 4000 3500uA 16U 2W 48.001 HEAD {* AC MICROAMP TESTS *} 48.002 JMP 49.001 48.003 EVAL 49.001 DISP Press the BLUE button to select AC. 49.002 HEAD AC MICROAMP TESTS: { 400uA Range } 49.003 5700 400 350.0uA 4.4U 60H 2W 50.001 5700 400 350.0uA 4.4U 1kH 2W 51.001 HEAD AC MICROAMP TESTS: { 4000uA Range } 51.002 5700 4000 3500uA 44U 60H 2W 52.001 5700 4000 3500uA 44U 1kH 2W 53.001 DISP Disconnect the 5700A from the UUT. 53.002 HEAD {* DC AMP TESTS *} 53.003 JMP 54.001 53.004 EVAL 54.001 DISP Rotate the UUT function switch to mA/A. 54.002 DISP Connect the 5725A and UUT as follows: 54.002 DISP [32] 5725A CURRENT OUTPUT HI to UUT A 54.002 DISP [32] 5725A CURRENT OUTPUT LO to UUT COM 54.003 HEAD DC AMP TESTS: { 4000mA Range } 54.004 5700 4000 3500mA 16U B1 2W 55.001 HEAD DC AMP TESTS: { 10A Range } 55.002 5700 10 10.00A 0.06U B1 2W 56.001 HEAD {* AC AMP TESTS *} 56.002 JMP 57.001 56.003 EVAL 57.001 DISP Press the BLUE button to select AC. 57.002 HEAD AC AMP TESTS: {* 4000mA Range *} 57.003 5700 4000 3500mA 44U 60H B1 2W 58.001 5700 4000 3500mA 44U 1kH B1 2W 59.001 HEAD AC AMP TESTS: { 10A Range } 59.002 5700 10 10.00A 0.14U 60H B1 2W 60.001 5700 10 10.00A 0.14U 1kH B1 2W 61.001 HEAD 61.002 END